          High-speed, High-accuracy CCD Laser Displacement SensorLK-G3000 series

          LK-G15

          Sensor Head: Super Precision, Wide beam

          SPECIFICATIONS

          Select Language

          Model

          LK-G15

          Reference distance

          10 mm

          Measuring range

          ±1 mm*1*2

          Light source

          Type

          Red semiconductor laser

          Wavelength

          650 nm (visible light)

          Laser Class

          Class II (FDA CDRH 21CFR Part 1040.10), Class 1 (IEC60825-1)

          Output

          0.3 mW

          Spot diameter (at reference distance)

          Approx. 20 x 500 µm

          Linearity

          ±0.03 % of F.S. (F.S.= ±1 mm)*3

          Repeatability

          0.02 µm (0.01 µm)*4

          Sampling frequency

          20 / 50 / 100 / 200 / 500 / 1000 µs (Selectable from 6 levels)

          LED display

          Near the centre of the measurement: Green lights
          Within the measurement area: Orange lights
          Outside the measurement area: Orange flashing

          Temperature characteristics

          0.01 % of F.S./°C (F.S.= ±1 mm)

          Environmental resistance

          Enclosure rating

          IP67 (IEC60529)

          Ambient light

          Incandescent lamp or Fluorescent lamp: 10,000 lux max.

          Ambient temperature

          0 to +50 °C

          Relative humidity

          35 to 85 % RH (No condensation)

          Vibration resistance

          10 to 55 Hz, Double amplitude 1.5 mm, 2 hours in each of the X, Y, and Z directions

          Material

          Aluminium die-cast

          Weight

          Approx. 190 g (including the cable)

          *1 The range is obtained by measuring KEYENCE's standard target (ceramic).
          *2 When the sampling rate is 20 µs, the range becomes +0.37 (FAR side) to -1 mm (NEAR side).
          *3 The range is obtained by measuring KEYENCE's standard target (ceramic) with the Standard mode.
          *4 The range is obtained by measuring KEYENCE's standard (SUS) with 4,096 times of averaging at the reference distance. The range in parenthesis is the typical linearity obtained by measuring the target with 16,384.